Portable TDL Laser Ammonia Analyzer EDK 6900P series


Overview

The EDK series products employ Enhanced TDLAS(Tunable Diode Laser Absorption Spectroscopy) technology with 0.1nm wavelength scanning, which mitigates cross-interference from absorption bands of interferent gases—also referred to as "fingerprint spectroscopy". This enables high-resolution near-infrared absorption measurements. Electron phase-locking technology facilitates the extraction and isolation of gas absorption signatures from photoelectric signals, eliminating the need for a physical reference cell while providing continuous sensor condition monitoring. The EDK TDL 6900 series offers a high-precision, low-detection-limit solution for in-situ gas measurement, featuring high accuracy, selectivity, zero calibration requirement, low cost, and ease of operation.

 

The EDK TDL 6900 series is capable of quantifying trace gases including NH, CH, CO, HO, etc., which are critical in various industrial applications. The inherent high sensitivity and wide dynamic range of TDLAS technology enable gas concentration measurements from sub-ppm levels to percentage levels.


Features


  • Exhibits Ultra - high Sensitivity

  • Functional Safety with Real-time Status Monitoring

  • Long Service Life (Over 10 Years)

  • Rapid Response

  • Low Power Consumption Design

  • Low-Cost Solution (No Consumables, Spare Parts, or Recalibration Required)

  • 19 - inch Mounting Bracket

  • Utilization of Optimal Cost - effective Components Ensures the Lowest - cost Sensor Design

  • User-friendly Human-Machine Interface Design

  • Robust Data Storage and Logging Functionality

  • Flexibly Scalable Multi - point Sampling System Architecture

  • Zero - point and Wide - range Calibration Function

  • Automatic Sensor Status Self - diagnostic Function


Measured   Gas

Minimum Detection Limit (MDL)

Typical   Measurement Range

NH3, (H2O) ammonia

(High-Temperature   & High-Humidity Working Conditions)

0.2   ppm

0~20,50,100,500,1000ppm,50%,the   range can be customized

NH3 ammonia

0.1   ppm

0~20,50,100,500,1000ppm,50%,the   range can be customized

Other   gases are customizable

**The   detection limit is at constant temperature, constant pressure, constant humidity   20°C, 1013 hPa, 50±1.5%r.H. Sudden changes in system   temperature result in faster changes in detection limits than changes in   concentration, and the whole process is 220°C with   high temperature tracing

accuracy

±2% of   Reading Value (RV) is based on integral stability (temperature &   pressure).

Span   Drift

Not   more than 2% per annum

Maximum   temperature error

Not   above 0.1 reading/°C

Linearity/Repeatability

<1%FS(multi-point   calibration),<2%FS(Single-point calibration)

Cell   Temperature

Constant   temperature 220°C (temperature can be set)

Display   Resolution

0.001   ppm

Refresh   Rate

Not   less than 1s (can be set, not less than 120s)

T90   Response Time

Not   higher than 1s (@chest capacity 3 L/min).

Ambient Temperature Compensation Range

-10~65°C   (smallest possible temperature fluctuation)

Maximum   Sample Gas Humidity

50% absolute humidity, depending on the   operating conditions, needs to be entered for calibration, and the pressure,   temperature or mixed gas concentration is compensated by diffusion

Maximum   Sample Volume

5L/min;   Calibration at 3 L/min

Power   Supply

VAC220-230/115   50/60 Hz

Size

51.85cmx40.5cmx23cm

 

Weight

Approx.   15kg

Enclosure

IP65   PP alloy suitcase

Explosion   Protection Rating

ExdII   CT4,ExdII CT5, (customizable)

Gas   Path Connection

Swagelok   Φ6 mm

Data   Storage

Built-in   no less than 8G, which can be expanded through USB port

Measurement   method

In-situ   extraction

Dust   removal filtration

Five-stage   filtration, filtration accuracy is not less than 0.2μm

Sampling   Pumps

Built-in   420°C negative pressure sampling pump

Analog   Output

4~20   mA Current Loop (No Isolation)

Port

Ethernet   port, USB port, Bluetooth, Wifi

Display

5.7"   touch screen

Interactive

  

It   can be connected to a mouse, keyboard and printer, and is not less than a   7-inch touch LCD screen

Sampling   Probes

 

0.5m,   1.5m long, 8mm~25mm outer diameter sampling probe greater than 220°C,   stainless steel or PTFE material, can be customized

Sampling   heat trace lines

5m   long 220°C heat tracing pipeline, outer diameter 6mm interface, with no   less than 25μm filtering function, power supply   220VAC, 50Hz, 200W, weight not more than 2.5kg